Purpose |
General radiography |
Method |
Scintillator + Amorphous Silicon (a-Si) sensor |
Sensor |
LANMIT2® (Large Area New-MIS sensor and TFT) |
Grid |
Attachable (Canon CXDI grid) – Choice of 4:1, 8:1, 10:1, horizontal and vertical |
Application |
Portable |
Pixel |
2,256 x 2,878 pixels (6,493,000 pixels) |
Image size |
Automatic sizing up to 9” x 11” (226 mm x 288 mm) |
Pixel pitch |
100 x 100 µ m |
A/D |
14-bit |
Grayscale |
4,096 (12-bit) grayscale |
Preview image |
Approx. 3 seconds after X-ray exposure (Actual times may vary due to various factors) |
Interval |
Approx. 6 seconds |
Total image processing |
Approx. 20 seconds per image (Actual times may vary due to various factors) |
Interface |
DICOM 3.0, Ethernet 10/100 Base T |
DICOM |
DICOM 3.0 compatible
|
Storage |
Temporary storage available in the Control PC |
Voltage |
100V, 120V, 230/240V (50Hz/60Hz) |
Power consumption |
300VA maximum |
Operation environment |
50-95°F (10-35°C)
30-75% RH (no condensation) |
Certification |
FDA510(k), FCC Class A, UL 2601-1, EN60601, CE 0197 |
Dimensions, weight
Sensor Unit |
12.8”W x 12.9”L x 0.8”H (324 x 327 x 20 mm), 6.2 lbs. (2.8kg) |
Operation/Preview Panel |
15.7”W x 15.5”L x 5.9”H (399 x 394 x 150 mm), 14.5 lbs. (6.6kg) |
Control PC |
11.8”W x 20.9”L x 23.4”H (300 x 532 x 594 mm) |
User Options
Grid |
Choice of 4:1, 8:1, 10:1, horizontal and vertical |
Data input |
Choices are:
-
DICOM Basic Modality Worklist Management Service Class (SCU
-
DICOM Modality Performed Procedure Step Service Class (SCU)
-
Other non-DICOM HIS/RIS communication
-
Barcode
For further details, please contact an authorized Canon Medical Systems representative. |