The RadPRO1 Mobile 40kW Flex Digital X-ray System, is a portable x-ray system that incorporates a new telescopic tube column helping radiographic technologists to maximize their lines of vision for safe transport when moving around a healthcare facility.  The new functionality builds on the ease of maneuvering and positioning that is a hallmark of the current RadPRO Mobile 40kW System, while maintaining the ability to capture high-quality diagnostic images for routine diagnosis or challenging trauma directly from a patients’ bed. 

Product Features

  • ▪ Telescopic tube column maximizes lines of vision for safe transport
  • ▪ New lead crystal batteries retain power longer, eliminating the need to recharge more frequently
  • ▪ Dual battery configuration allows for distributed power to the drive motors, as well as the generator
  • ▪ Secondary drive controls on the collimator handles allow the system to be positioned without the need to return to the main console
  • ▪ A patent-pending wireless distributed antennae system that can increase wireless range 40% over single access point systems
  • ▪ Integrates with choice of Canon CXDI 401C/701C/801C wireless flat panel detectors 

Quick Specs

  • 40 kW x-ray generator (40-150 kVp) 
  • 300 khu x-ray tube 
  •  0.7,1.3mm focal spots
  • 16 degree target angle (important for being able to take images at shorter SID without adverse heel-effect)  
  • Faster Battery charge time (11% minimum)
  • Additional spectral filtration in collimator
  • Laser light-field indicator 

Product Specifications

  • Counterbalanced Tube Arm:
  • Tube arm reach 49.2 inch (1,250 mm) Column rotation: ± 315°
  • Maximal SID to floor 80.7 inch (2,050 mm)
  • Minimal SID to floor 21.6 inch (550 mm)
  • MOBILE UNIT:
  • Weight 1,268 lbs (575 kg) standard unit 1,312 lbs (595 kg) unit with telescopic column
  • Size (L x W x H) 53.54 x 26.4 x 77.9 inch (1,360 x 670 x 1,980 mm)
  • Touch screen console 17 inch (430 mm) diagonal
  • Anti-collision system Frontal
  • Motorized unit Up to 4 km/h
  • Motor Drive (Standard) Two (2) independent drive motors, one for each wheel (forward and reverse)
  • GENERATOR SPECS:
  • kW Rating 40 kW
  • kVp Range 40 to 150 kVp
  • mA Range 10 - 500 mA Selectable
  • Number of Stations 18
  • Exposure Time Range 0.001 - 10 sec.
  • mAs Range 0.1 - 500 mAs
  • Generator Integration Yes, 3 PT technique
  • Battery Capacitor Battery / Capacitor
  • Anatomical Programs 500+ Anatomical settings
  • Line Requirements Single Phase Line Regulation from 100 - 240 Vac (+/-10%)

RadPRO FLEX Data Sheet PDF (pdf)

Related Products:

CORE Medical Imaging - A Shared Vision for Better Health. Contact Us